    Inductively coupled plasma emission

    ICP-6810 in the nonferrous gold industry

    Time:2022-06-30  Read:

    ICP6810 in the nonferrous gold industry

    The ICP6810 profile is introduced

    The ICP6810 full-spectrum direct-read inductance-coupled plasma emission spectrometer, with excellent resolution, test accuracy and precision, is widely used in the rare earth industry, chemical industry, andOre analysis, metal smelting, geological research, drug safety, experimental research, environmental testing, food safety and other industries.Especially for the determination of elements in the ICP6810 zinc oxide industry.

    picture1.ICP6810

    Table 1. Main technical parameters of ICP6810

    ICP6810 main technical parameters

    High frequency ionizer

    Working frequency

    27.12MHz

    Frequency stability

    0.05

    Output power

     800W 1600W

    Output power stability

     0.05

    OMatching way

    Automatic matching

    Scan the light divider

    light path

    Full reflection imaging light path, quartz prism two-dimensional dispersion system

    focal distance

    400mm

    Graster specifications

    Middle step grating, 52.67, l p / mm, 64 shining angle

    Resolution ratio

    <0.0068nm@200nm

    Complete machine technical indicators

    Scan the wavelength range

    At 165-900nm, full wavelength coverage, and measurable Al 167.120nm

    Repetitiveness

    RSD1.5

    Stability

    RSD2.0

    ICP6810 element detection limit (g / L)

    element

    La

    Ce

    Pr

    Nd

    Sm

    Eu

    Gd

    Tb

    wavelength (nm)

    408.672

    413.765

    414.311

    401.225

    360.946

    381.967

    342.247

    350.917

    detection limit

    <3.0

    <5.0

    <5.0

    <5.0

    <10.0

    <1.0

    <10.0

    <3.0


    element

    Dy

    Ho

    Er

    Tm

    Yb

    Lu

    Y

    Sc

    wavelength (nm)

    353.170

    345.600

    337.271

    313.126

    369.419

    261.541

    371.030

    335.373

    detection limit

    <3.0

    <3.0

    <3.0

    <3.0

    <1.0

    <3.0

    <1.0

    <1.0


    element

    Ta

    Nb

    Mn

    Mg

    B

    Zn

    Co

    Si

    wavelength (nm)

    226.230

    313.340

    257.610

    279.553

    249.773

    213.856

    228.616

    251.611

    detection limit

    <5.0

    <5.0

    <3.0

    <1.0

    <10.0

    <3.0

    <3.0

    <10.0


    element

    Ni

    Cd

    Fe

    Ca

    Mo

    V

    Be

    Ti

    wavelength (nm)

    232.003

    226.502

    239.562

    393.366

    281.615

    310.230

    313.041

    334.941

    detection limit

    <5.0

    <3.0

    <3.0

    <1.0

    <5.0

    <5.0

    <1.0

    <3.0


    element

    Cu

    Cr

    Al

    Zr

    Ag

    Sr

    Au

    Pt

    wavelength (nm)

    324.754

    267.716

    396.152

    343.823

    328.068

    407.771

    242.795

    265.945

    detection limit

    <3.0

    <5.0

    <5.0

    <5.0

    <3.0

    <1.0

    <5.0

    <5.0


    element

    Pd

    Ir

    Rh

    Ru

    Ba

    Li

    Na

    K

    wavelength (nm)

    340.458

    224.268

    343.489

    240.272

    455.403

    670.784

    588.995

    766.490

    detection limit

    <5.0

    <10.0

    <10.0

    <5.0

    <1.0

    3

    20

    60


    element

    As

    Sb

    Bi

    Hg

    Pb

    Ga

    Os

    W

    wavelength (nm)

    228.812

    206.833

    223.061

    253.652

    220.353

    294.364

    225.585

    207.911

    detection limit

    15

    15

    10

    15

    15

    10

    1

    10


    element

    Sn

    Te

    Ta

    Th

    Tl

    Re

    Ge

    Se

    wavelength (nm)

    242.949

    214.281

    226.230

    283.730

    276.787

    227.525

    209.426

    203.985

    detection limit

    20

    10

    5.0

    10

    30

    5

    15

    30

    evalsuation of alloy measurable elements in non-ferrous precious metals                                                                                                                                               

    For precious metal samples and customer test requirements, the details are shown in the Table:

    2. Precious metal sample test solution

    sample

    measureelement

    Analysis instrument

    Sample handling overview (for reference only)

    remarks

    Non-ferrous precious metal alloy

    Gold, silver, palladium, platinum, and so on

    ICP6810

    Take 0.1-0.3g sample in a 100ml beaker, add 5ml hydrochloric acid, dissolve the sample on the heating plate to clarification, remove cold to room temperature, transferred to 50ml volumetric flask diluted with water to scale for testing.

    Sample blank experiment was also performed simultaneously duringthesampleprocessing process.


    According to the configured standard solution concentration, directly on the computer test, the instrument can automatically give the final test results, and can be directly printed in the form of a report.

    Table 3.ICP6810 and related instrument configuration

    Instrument host and necessary accessories

    order number

    definition

    Specifications / models

    manufacturer

    quantity remarks

    quantity remarks

    1

    Full-spectrum straight-read inductively coupled plasma emission spectrometer

    ICP6810

    Shanghai Beauty Instrument Co., Ltd

    1

     

    2

    Automatic temperature control and circulating cooling water device

    1

     

    3

    stabilized voltage supply

    1

     

    Instrument matching accessories

    order number

    definition

    Specifications / models

    manufacturer

    quantity remarks

    quantity remarks

    1

    host

    1

     

    2

    graphoscope

    1

     

    3

    marking punch

    1

     

    Sample processing equipment and consumables

    order number

    definition

    Specifications / models

    manufacturer

    quantity remarks

    quantity remarks

    1

    Electronictemperature-controlled heating board

    EG-35B

    1

     

    2

    One in ten thousand is an electronic balance

    BS224S

    1

     

    3

     Laboratory-grade ultra-pure water machine

    EPED-40T

    1

     

    4

    Disposable rubber Gloves

    M

     

    /

     

    5

    Activated carbon mask

     

     

    /

     

    6

    3mlpasteurpipette

    3ml

     

    /

     

    7

    100mlglass beaker

    100ml

    Sichuan Shu Glass scientific Instruments

    /

     

    8

    100mlGlass capacity bottle

    100ml

    Sichuan Shu Glass scientific Instruments

    /

     

    9

    5mltransfer liquid gun

    0.5ml-5ml

    Rand, Germany

    1

     

    10

    1mltransfer liquid gun

    100ul-1000ul

    Rand, Germany

    1

     

    Reagent

    order number

    definition

    Specifications / models

    manufacturer

    quantity remarks

    quantity remarks

    1

    hydrochloric acid GR Jiangsu Qiangsheng Chemical Co., Ltd

    hydrochloric acid GR Jiangsu Qiangsheng Chemical Co., Ltd

    hydrochloric acid GR Jiangsu Qiangsheng Chemical Co., Ltd

    /

     

    2

    hydrogen nitrate GR Jiangsu Qiangsheng Chemical Co., Ltd

    hydrogen nitrate GR Jiangsu Qiangsheng Chemical Co., Ltd

    hydrogen nitrate GR Jiangsu Qiangsheng Chemical Co., Ltd

    /

     

    3

    perhydrol GR Jiangsu Qiangsheng Chemical Co., Ltd

    perhydrol GR Jiangsu Qiangsheng Chemical Co., Ltd

    perhydrol GR Jiangsu Qiangsheng Chemical Co., Ltd

    /

    4

    Standard material of each element 1000ug/ml National nonferrous Metals and electronic materials center of forecasting and analysis

    Standard material of each element 1000ug/ml National nonferrous Metals and electronic materials center of forecasting and analysis

    Standard material of each element 1000ug/ml National nonferrous Metals and electronic materials center of forecasting and analysis

    /

     


    If you have any other questions, please call the analysis instrument